Human Epstein-barr virus capsid antigen IgG antibody (EBV-VCA-IgG) ELISA Kit

Citations(0) Uniprot : N/A
  • Cat.No.: AE64799HU

  • Reactivity: Human (Homo sapiens)

To Purchase AE64799HU

Size:
48T 96T
Price: 查询

Reagent Preparation







Results demonstration







Assay Procedure Summary







Certificate



Product Details

种属 Human (Homo sapiens)
UniProt号 N/A
缩写 EBV-VCA-IgG
别名 N/A
线性范围 Request Information
灵敏度 Request Information
样本类型 Serum, Plasma, Other biological fluids
检测方法 Competitive
分析方法 Qualitative
反应时间 1-3h
样本体积 1-200 μL
检测波长 450 nm

原理

This assay employs a two-site sandwich ELISA to quantitate EBV-VCA-IgG in samples. An antibody specific for EBV-VCA-IgG has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any EBV-VCA-IgG present is bound by the immobilized antibody. After removing any unbound substances, a biotin-conjugated antibody specific for EBV-VCA-IgG is added to the wells. After washing, Streptavidin conjugated Horseradish Peroxidase (HRP) is added to the wells. Following a wash to remove any unbound avidin-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of EBV-VCA-IgG bound in the initial step. The color development is stopped and the intensity of the color is measured.
 

产品概述


 

组份


Reagents

Quantity

Reagents

Quantity

Assay plate (96 Wells)

1

Instruction manual

1

Standard (lyophilized)

2

Sample Diluent

1 x 20 mL

Biotin-Conjugate (concentrate 100 x)

1 x 120 μL

Biotin-Conjugate Diluent

1 x 12 mL

Streptavidin-HRP  (concentrate 100 x)

1 x 120 μL

Streptavidin-HRP Diluent   

1 x 12 mL

Wash Buffer (concentrate 25 x)

1 x 20 mL

Substrate Solution

1 x 10 mL

Stop Solution

1 x 6 mL

Adhesive Films

4

 

特异性

This assay has high sensitivity and excellent specificity for detection of Human EBV-VCA-IgG. No significant cross-reactivity or interference between Human EBV-VCA-IgG and analogues was observed.
 

回收率

Matrices listed below were spiked with certain level of recombinant Human EBV-VCA-IgG and the recovery rates were calculated by comparing the measured value to the expected amount of Human EBV-VCA-IgG in samples.

Sample Type

Number

Recovery range (%)

Average(%)

Serum

10

90-101

96

EDTA plasma

10

89-97

93

Heparin plasma

10

91-99

95

 

精密度

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. CV (%) = SD/meanX100 Intra-Assay: CV<8% Inter-Assay: CV<13%
 

线性


Sample Type

1:2

1:4

1:8

1:16

Serum

78-89%

81-99%

92-103%

95-105%

EDTA plasma

91-101%

90-98%

93-101%

91-98%

Heparin plasma

92-103%

93-102%

92-99%

91-101%

 

稳定性


 

样本收集及储存

Serum: Use a serum separator tube (SST) and allow samples to clot for two hours at room temperature or overnight at 2 - 8°C before centrifugation for 15 minutes at 1000 × g. Remove serum and assay immediately or aliquot and store samples at ≤ -20°C. Avoid repeated freeze-thaw cycles. Plasma: Collect plasma using EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 × g at 2 - 8°C within 30 minutes of collection. Assay immediately or aliquot and store samples at ≤ -20°C. Avoid repeated freeze-thaw cycles. Other biological fluids: Centrifuge samples for 20 minutes at 1000 × g. Remove particulates and assay immediately or store samples in aliquot at -20°C or -80°C. Avoid repeated freeze/thaw cycles.
 

注意事项

Store at 2-8°C. Please refer to Instruction Manual.